Triplett 9007 Manual do Utilizador Página 37

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 48
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 36
Page 36
7.10.5 If the displayed value is less than “200” (decimal point not
shown), a more accurate reading may be obtained by setting
the RANGE switch to a lower range. Rotate the RANGE switch
to the next lower position and observe the reading on the LCD
display. If the RANGE switch is already on the lowest position
(i.e. 2n F), no greater measurement resolution can be obtained.
Notes:
a)Onthe2nFand20nFranges,themetermayindicateresidual
capacitance.i.eThemeterreadingwillnotgotozero.Thisis
causedbytheinternalcapacitanceofthemetercircuitryorthe
testleads.
b)TheTestLeadsmayintroduceerrorwhenusedonthelow
capacitanceranges(like2nFand20nF).Tominimizetheerror,
notethedifferenceinreadingwiththeleadsinsertedintothe
meterandwiththeleadsremovedfromthemeter,andwhen
usingthetestleads,subtractthenoteddifferencefromthe
displayedreading.
Vista de página 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 47 48

Comentários a estes Manuais

Sem comentários